INSTALLING THE CHART PAPER

The recorder accepts three different types of chart paper. Charts are available for 1, 7 and 32 day recording times. Although chart paper replacing is very simple, you must take care to set the appropriate front panel switches to match the paper being used. Set the chart type and Cal/Normal switches (refer to Figure 2-7) to the following positions for these three (3) modes of operation. Figure 2-1 shows the removal of the decorative foot cover.

Chart Type Switch (Days)
CAL/NORMAL
Paper Type
1
C
1-day, Celsius and pH paper
7
C
7-day, Celsius and pH paper
32
C
32-day, Celsius and pH

Chart Switches
Figure 2-7. Chart Switches

Stabilizing Arm Fully Extended
Figure 2-8. Stabilizing Arm Fully Extended
To change the chart paper:
1a. Table Mounted Recorder

To balance the recorder, fully extend the stabilizing arm at the bottom of the recorder. Refer to Figure 2-8. Continue to Step 2.


1b. Wall Mounted Recorder:
You can also leave the recorder on the wall if it is easily accessible.

2. Open the door and hold it open with left hand.

Changing Chart paper and Setting the Time
Figure 2-9
Changing Chart Paper and Setting the Time
3. Remove the magnetic chart paper knob (Figure 2-9, item #2) and used chart paper.

4. Place the new chart paper (Figure 2-9, item #1) on the spindle and rotate it until the current time on the chart paper is aligned with the time arrow on the chart base of the recorder. See Figure 2-9 detail.

5. When the chart paper is correctly oriented, replace the magnetic knob (Figure 2-9, item #2), being careful not to alter the position of the chart paper.

6. Make sure that the ôChart Typeö switch, located behind the control panel door, is in the correct position relative to the chart paper used.


NOTE: If the switch is in the wrong position, the chart paper will not rotate at the correct speed and the indicated chart time will not correspond to real time.

Installing and Removing Pens
The recorder uses two different colored pens to record information. The green (upper) pen records pH. The red (lower) pen records temperature. Pen life varies with use and climate. Under normal operating conditions, pens last at least one month. We recommend that you install a fresh set of pens after every month of use.

1. Make sure the power switch on the front panel is in ôOö position. The front panel is located under the control panel door. The recorder door should be closed.

NOTE: The control panel door opens easily by gently gripping the recessed area at the right side and slowly pulling down.

2. Examine the location of the pen arms through the window in the recorder door.

a. If the pen arms are offset as shown in Figure 2-10,continue to Step 3:


Figure 2-10. Offset Alignment of Pens
(for ease of pen installation)
b. If the pen arms are vertically aligned as shown in Figure 2-11, do the following and then proceed to Step 3.

Figure 2-11. Vertical Alignment of Pens
(pens canÆt be inserted)

(The pen arms must be offset sufficiently to allow for clearance before insertion or removal is attempted. The pens must be moved electrically.)

  • Make sure the recorder door is closed and a piece of chart paper is installed on the spindle on the chart base.
  • Turn the unit back on by moving the power switch to ô|ö and wait for the pens to be far enough apart to remove a used pen one at a time or to insert a new pen one at a time.


CAUTION: DO NOT ATTEMPT TO OFFSET THE PENS MANUALLY (BY HAND). THE DRIVE MECHANISM COULD BE DAMAGED. DO NOT ATTEMPT TO CHANGE THE PENS WHILE THEY ARE IN MOTION.

  • Turn off the unit (PWR switch position in ôOö position) to stop the pen movement.


3. Open the recorder door by applying downward pressure on the latch button (refer to Figure 1-1, item #16).
4. Remove the pens from the package and the cap from each pen.

We recommend that you put the pen caps on special posts located on the chart base of the recorder near the light bulbs.Figure 2-12 shows the location of the posts.


Post Location for Pen Caps
Figure 2-12. Post Location for Pen Caps
Pen Installation
Figure 2-13. Pen Installation

5. Insert the red pen fully into the lower (temperature) holder and the green pen fully into the upper (pH recorder) holder. If you installed the pens correctly, you will hear a click as they position themselves in the holders. Figure 2-13 shows how to insert the pens.
6. Close the recorder door.

Removing the Pens
1. Check to see that the pens are offset as shown in Figure 2-10. Otherwise, you can damage the arm and drive mechanism if you try to remove them.

CAUTION: DO NOT ATTEMPT TO OFFSET THE PENS MANUALLY (BY HAND). THE DRIVE MECHANISM COULD BE DAMAGED. DO NOT ATTEMPT TO CHANGE THE PENS WHILE THEY ARE IN MOTION.
Pen Removal
Figure 2-14. Pen Removal
2. Perform Step 2b in Section 2.5.1 to offset the pens.

3. With a screwdriver in one hand, push on the rectangular area of the pen shown in Figure 2-14. Using the other hand, release the pen from the spring clip attached to the pen holder.

4. At the same time, slide the pen out from the pen holder.
5. Install new pens following Section 2.5.1, Steps 3

Installing the pH/Temperature Sensor
The pH/temperature sensor can either be directly attached to the recorder using its own attached short cable or be extended by using the remote sensor cable.

Using the Short Cable
1. Place the handle part of the sensor in to the sensor clip on the side of the recor
der.
2. Install the connector into the socket at the side of the recorder. Make sure the sensor cable is untwisted.


Using the Remote Sensor Cable

1. Connect the end of the cable with the arrow to the short cable on the sensor. Line up the two arrows as shown in Figure 2-16 before mating the two cables.

2. Connect the other end of the remote sensor cable (without the arrow) to the socket at the side of the recorder.


Figure 2-16.(On right)
Using the Remote Sensor Cable
Using the Remote Sensor Cable
Help Price List Specifications Print Manual Manuals Home
 Home Home  Table of Contents TOC  Back Back  Next Next